BRUKER WB 500
in operation since June 2016
This spectrometer is mainly used in solid-state mode. If required, it can be converted to high resolution.
Probe: PA TBO 500 W2/S4 BB H&F/P-D-05 Z
Triple configuration
2H coil (Lock)
1H and 19F are located on the same coil.
Additional 31P coil
Heteronuclei (X) coil
1H and 19F are located on the same coil:
→ Heteronuclei (X) can be measured with 1H-decoupling OR 19F-decoupling.
→ 1H heteronucleus (X) 2D measurements are possible (e.g. 1H 13C HMQC).
→ 19F heteronucleus (X) 2D measurements are possible (e.g. 19F 13C HMQC).
→ 19F CANNOT be measured with 1H decoupling.
→ 1H CANNOT be measured wtih 19F decoupling.
This probe is equipped with an additional phosphorus coil (triple configuration).
→ Heteronuclei (X) can be measured with 31P decoupling in addition to 1H OR 19F.
→ 31P heteronucleus 2D measurements are possible (e.g. 31P 13C HMQC) with or without further 1H OR 19F decoupling
(e.g. 31P 13C{19F} HMQC).
Temperature experiments can be carried out within a range from -40°C up to +150°C. Care must be taken to select a suitable solvent.
For temperature measurements and experiments not carried out automatically, please make an appointment.
Experiments:
Further experiments in progress or on request
| 1H | 1H | ||
| 1H1H COSY | HH correlation within one spin system | ||
| 11B | 11B | with and without background suppression | |
| 11B{1H} | with and without background suppression | ||
| 1H11B HMQC | |||
| 1H11B HMBC | |||
| 13C | 13C{1H} | ||
| 13C{1H} dept135 | CH3/CH-positive – CH2-negative | ||
| 13C{1H} jmod = apt | CH3/CH positive – CH2/C negative | ||
| 1H13C HMQC | |||
| 1H13C HMBC | |||
| 13C19F | 13C{19F} | 19F decoupled BUT 1H coupled | |
| For further 13C19F experiments see at 19F | |||
| 13C31P | 13C{31P1H} | with 31P decoupling AND 1H decoupling | |
| 13C{31P} | with 31P decoupling BUT 1H coupled | ||
| For further 13C31P experiments see at 31P | |||
| 15N | ¹H¹⁵N HMQC | ||
| 1H15N HMBC | |||
| 15N{1H} inept-rd | Polarisation transfer | ||
| 15N inept-pnd | Polarisation transfer | ||
| 19F | 19F | Only 1H coupled possible | |
| 19F19F COSY | |||
| ¹H¹⁹F | 1H19F COSY | ||
| 19F11B | 19F11B HMBC |
→ Coupling constants have to be specified.
| |
| 19F13C | 19F13C HMQC | ||
| 19F29Si | 19F29Si HMBC | ||
| 19F77Se | 19F77Se HMQC | ||
| 19F77Se HMBC | |||
| 19F125Te | 19F125Te HMBC | ||
| 27Al | 27Al | with and without background suppression | |
| 27Al{1H} | with and without background suppression | ||
| 1H27Al HMBC | |||
| 29Si | 29Si{1H} zgig | inverse gated | |
| 29Si{1H} inept-rd | Polarisation transfer | ||
| 29Si inept-pnd | Polarisation transfer | ||
| 1H29Si HMQC | |||
| 1H29Si HMBC | |||
| 29Si19F | 29Si{19F} zgig | inverse gated – with 19F decoupling BUT 1H coupled | |
| 29Si{19F} inept-rd | Polarisation transfer – with 19F decoupling BUT 1H coupled | ||
| 31P | 31P{1H} | ||
| 1H31P HMQC | |||
| 1H31P HMBC | |||
| 31P13C | 31P{13C1H} | with 13C decoupling AND 1H decoupling | |
| 31P13C HSQC |
→ Coupling constants have to be specified | ||
| 31P13C HMBC | |||
| 31P15N | 31P15N HSQC | ||
| 31P29Si | 31P29Si HSQC | ||
| 31P29Si HMBC | |||
| 31P{29Si1H} | with 29Si decoupling AND 1H decoupling | ||
| 77Se | 77Se{1H} | ||
| 1H77Se HMQC | |||
| 1H77Se HMBC | |||
| 77Se19F | 77Se{19F} | with 19F decoupling BUT 1H coupled | |
| 119Sn | 119Sn{1H} zgig | inverse gated | |
| 1H119Sn HMQC | |||
| 1H119Sn HMBC | |||
| 125Te | 125Te{1H} zgig | inverse gated | |
| 1H125Te HMQC | |||
| 1H125Te HMBC | |||