JEOL JNM-ECZL500R
in operation since May 2023
Two sample heads are available for this NMR spectrometer. The Supercool sample head is fitted as standard.
Supercool sample head: NM-83020SCO5TH5 500MHZ 5MM FG/SCOTH
Cooling with liquid nitrogen
Equipment: ¹H/¹⁹F coil
X coil

- ¹H and ¹⁹F are located on one coil:
- Hetero-cores (X) can be measured¹H-decoupled OR ¹⁹F-decoupled.
- ¹H hetero-core 2D measurements are possible. (e.g.¹H¹³C HMQC)
- ¹⁹F Hetero-core 2D measurements are possible. (e.g. ¹⁹F¹³C HMQC)
- ¹⁹F can NOT be measured ¹H-decoupled.
- ¹H can NOT be measured ¹⁹F decoupled.
The Supercool probe head is cooled down to approx. -196°C with liquid nitrogen. This reduces the noise caused by the electronics. This electronic noise is temperature-dependent. This improves sensitivity and reduces measurement time. The reduction in measuring time depends on the core and the experiment. The sample chamber remains at room temperature or the desired measuring temperature.
Temperature experiments are possible from room temperature to +150°C. Care must be taken to select a suitable solvent.
Sample head ROYAL HFX: NM-80010HFXSS 500MHZ FG/ROHFX
Equipment: ¹H/¹⁹F coil with DUAL MODE
X coil

This sample head is equipped with an additional dual tune mode for 1H and 19F. Although the two cores are on one coil, they can be measured simultaneously. I.e.
- ¹H can be measured ¹⁹F decoupled AND vice versa.
- Hetero-cores can be measured simultaneously ¹H- and ¹⁹F-decoupled or either only ¹H-decoupled or only ¹⁹F-decoupled.
- ¹⁹F hetero-core 2D measurements are possible (e.g. ¹⁹F¹³C HMQC) with or without additional 1H decoupling.
¹H Hetero-core 2D measurements are possible (e.g. ¹H¹³C HMQC) with or without additional 19F decoupling.
Temperature experiments are possible from -100°C to +150°C. Care must be taken to select a suitable solvent.
__________________________________________________________________________________
For temperature measurements and experiments that are not measured in the automatic system, please make an appointment.
Experiments - installed so far:
Further experiments in progress or on request.
| ¹H: | ¹H | |
| ¹H_para | ||
| ¹H¹H COSY | H-H correlation within a spin system | |
| ¹H¹H NOESY | ||
| NOE diff. | Differential measurement | |
| ¹H{¹⁹F} | HFX sample head only | |
| ¹H{³¹P} | Decoupling also possible with 2 offsets | |
| ¹H{X} | Decoupling from other heteronuclei is possible; also possible with 2 offsets | |
| PSYCHE | ¹H{¹H} | |
| DOSY | diffusion | |
| EXSY | Chemical exchange | |
| Water suppression | ||
| Solvent suppression | For non-deuterated solvents / measurement without lock | |
| ²H | Deuterium | Measurement on the lock channel |
| ⁷Li: | 7Li{1H} | |
| ¹¹B: | 11B | with and without background suppression |
| 11B{1H} | with and without background suppression | |
| 1H11B HMQC | ||
| 1H11B HMBC | ||
| ¹³C: | ¹³C{¹H} | |
| ¹³C{¹H} dept135 | CH₃/CH-positive - CH₂-negative | |
| ¹³C{¹H} apt | CH₃/CH-positive - CH₂/C-negative | |
| ¹H¹³C HMQC | not phase sensitive | |
| ¹H¹³C HMBC | ||
| ¹⁵N: | ¹H¹⁵N HMQC | Coupling constant + measuring range must be specified |
| ¹H¹⁵N HMBC | Coupling constant + measuring range must be specified | |
| ¹⁹F: | ¹⁹F | |
| ¹⁹F{¹H} | HFX sample head only | |
| ²⁷Al: | ²⁷Al | with and without background suppression |
| ²⁷Al{¹H} | with and without background suppression | |
| ²⁹Si: | ²⁹Si | |
| ²⁹Si{¹H} | ||
| ²⁹Si{¹H}-inept | Polarisation transfer | |
| ²⁹Si-inept | Polarisation transfer | |
| ¹H²⁹Si HMQC | ||
| ¹H²⁹Si HMBC | ||
| ³¹P: | ³¹P | |
| ³¹P{¹H} | ||
| ¹H³¹P HMQC | ||
| ¹H³¹P HMBC | ||
| ⁷¹Ga: | ⁷¹Ga{¹H} | |
| ⁷⁷Se: | ⁷⁷Se | |
| ⁷⁷Se{¹H} | ||
| ¹H⁷⁷Se HMQC | ||
| ¹H⁷⁷Se HMBC | ||
| ¹¹⁹Sn: | ¹¹⁹Sn{¹H} | |
| ¹H119Sn HMBC | ||
| ¹²⁵Te: | ¹²⁵Te{¹H} | |
| ¹H¹²⁵Te HMBC | ||
| ¹⁹⁵Pt | ¹⁹⁵Pt{¹H} | |
| ²⁰⁷Pb: | 207Pb{¹H} | |