JEOL JNM-ECZL500R

in operation since May 2023

 

Two sample heads are available for this NMR spectrometer. The Supercool sample head is fitted as standard.

Supercool sample head: NM-83020SCO5TH5 500MHZ 5MM FG/SCOTH
Cooling with liquid nitrogen

Equipment: ¹H/¹⁹F coil
X coil

  • ¹H and ¹⁹F are located on one coil:
  • Hetero-cores (X) can be measured¹H-decoupled OR ¹⁹F-decoupled.
  • ¹H hetero-core 2D measurements are possible. (e.g.¹H¹³C HMQC)
  • ¹⁹F Hetero-core 2D measurements are possible. (e.g. ¹⁹F¹³C HMQC)
  • ¹⁹F can NOT be measured ¹H-decoupled.
  • ¹H can NOT be measured ¹⁹F decoupled.

The Supercool probe head is cooled down to approx. -196°C with liquid nitrogen. This reduces the noise caused by the electronics. This electronic noise is temperature-dependent. This improves sensitivity and reduces measurement time. The reduction in measuring time depends on the core and the experiment. The sample chamber remains at room temperature or the desired measuring temperature.

Temperature experiments are possible from room temperature to +150°C. Care must be taken to select a suitable solvent.

 

Sample head ROYAL HFX: NM-80010HFXSS 500MHZ FG/ROHFX

Equipment: ¹H/¹⁹F coil with DUAL MODE
X coil

This sample head is equipped with an additional dual tune mode for 1H and 19F. Although the two cores are on one coil, they can be measured simultaneously. I.e.

  • ¹H can be measured ¹⁹F decoupled AND vice versa.
  • Hetero-cores can be measured simultaneously ¹H- and ¹⁹F-decoupled or either only ¹H-decoupled or only ¹⁹F-decoupled.
  • ¹⁹F hetero-core 2D measurements are possible (e.g. ¹⁹F¹³C HMQC) with or without additional 1H decoupling.
    ¹H Hetero-core 2D measurements are possible (e.g. ¹H¹³C HMQC) with or without additional 19F decoupling.

Temperature experiments are possible from -100°C to +150°C. Care must be taken to select a suitable solvent.
__________________________________________________________________________________

For temperature measurements and experiments that are not measured in the automatic system, please make an appointment.

Experiments - installed so far:
Further experiments in progress or on request.

¹H:¹H
¹H_para
¹H¹H COSYH-H correlation within a spin system
¹H¹H NOESY
NOE diff.Differential measurement
¹H{¹⁹F} HFX sample head only
¹H{³¹P}Decoupling also possible with 2 offsets
¹H{X}Decoupling from other heteronuclei is possible; also possible with 2 offsets
PSYCHE¹H{¹H}
DOSYdiffusion
EXSYChemical exchange
Water suppression
Solvent suppressionFor non-deuterated solvents / measurement without lock
²HDeuteriumMeasurement on the lock channel
⁷Li:7Li{1H}
¹¹B:11Bwith and without background suppression
11B{1H} with and without background suppression
1H11B HMQC
1H11B HMBC
¹³C:¹³C{¹H}
¹³C{¹H} dept135CH₃/CH-positive - CH₂-negative
¹³C{¹H} aptCH₃/CH-positive - CH₂/C-negative
¹H¹³C HMQCnot phase sensitive
¹H¹³C HMBC
¹⁵N:¹H¹⁵N HMQCCoupling constant + measuring range must be specified
¹H¹⁵N HMBCCoupling constant + measuring range must be specified
¹⁹F:¹⁹F
¹⁹F{¹H}HFX sample head only
²⁷Al:²⁷Alwith and without background suppression
²⁷Al{¹H} with and without background suppression
²⁹Si:²⁹Si
²⁹Si{¹H}
²⁹Si{¹H}-ineptPolarisation transfer
²⁹Si-ineptPolarisation transfer
¹H²⁹Si HMQC
¹H²⁹Si HMBC
³¹P:³¹P
³¹P{¹H}
¹H³¹P HMQC
¹H³¹P HMBC
⁷¹Ga:⁷¹Ga{¹H}
⁷⁷Se:⁷⁷Se
⁷⁷Se{¹H}
¹H⁷⁷Se HMQC
¹H⁷⁷Se HMBC
¹¹⁹Sn:¹¹⁹Sn{¹H}
¹H119Sn HMBC
¹²⁵Te:¹²⁵Te{¹H}
¹H¹²⁵Te HMBC
¹⁹⁵Pt¹⁹⁵Pt{¹H}
²⁰⁷Pb:207Pb{¹H}

 